Abstract

Introduction: Ozonization is an alternative sterilization process for heat-sensitive medical devices. However, the side effects of this process on packaging materials should be verified. Methods: Four types of commercial disposable packaging for medical devices were evaluated after undergoing ozone sterilization: crepe paper sheet, non-woven fabric sheet (SMS), medical grade paper-plastic pouch and Tyvec©-plastic pouch. For each material, the gas penetration through the microbiological barrier was measured. Other packaging properties, such as chemical composition, color, tactile and mechanical resistance, were also evaluated after sterilization, by using characterization techniques, namely microbiological indicators, infrared spectroscopy, tensile test and optical microscopy. Results: All commercial disposable packaging showed good ozone penetration. Crepe paper and SMS were chemically and mechanically modified by ozone, while Tyvec© only suffered mechanical modification. Paper-plastic pouch was the packaging material which just experienced an acceptable reduction in tensile resistance, showing no variations on chemical or visual properties. Conclusion: The results suggest that medical grade paper-plastic pouch is the most appropriate disposable medical device packaging to be sterilized by ozone when compared to other materials.    

Keywords

Ozone, Packaging, Medical device, Sterilizer, FTIR, Tensile test.    

References

Ackert-Burr C. Low-temperature sterilization: are you in the know? Perioperative Nursing Clinics. 2010; 5(3):281-90. http://dx.doi.org/10.1016/j.cpen.2010.05.002.

American Society for Testing and Materials. Designation D882-02: standard test method for tensile properties of thin plastic sheeting. West Conshohocken: ASTM; 2002. 

Batakliev T, Georgiev V, Anachkov M, Rakovsky S, Zaikov GE. Ozone decomposition. Interdisciplinary Toxicology. 2014; 7(2):47-59. PMid:26109880. http://dx.doi.org/10.2478/intox-2014-0008. 

Billings S. Medical packaging design. Design Week. 2005; 20(44):8. 

British Standards Institution. BS EN 868-5:2009: packaging for terminally sterilized medical devices: sealable pouches and reels of porous and plastic film construction: requirements and test methods. London: BSI; 2009. 

Burlingame B, Denholm B, Link T, Ogg MJ, Spruce L, Spry C, Wood SAVWA. Guideline for selection and use of packaging systems for sterilization. Guidelines for Perioperative Practice AORN. 2016; 1:1-6. http://dx.doi.org/10.6015/psrp.15.01.651. 

DuPont. Dupont technical reference guide for medical and pharmaceutical packaging. Wilmington; 2011. 44 p. 

Ebihara K, Stryczewska HD, Mitsugi F, Ikegami T, Sakai T, Pawlat J, Teii S. Recent development of ozone treatment for agricultural soil sterilization and biomedical prevention. Przegl Elektrotech. 2012; 88(6):92-4. 

Eto H, Ono Y, Ogino A, Nagatsu M. Low-temperature sterilization of wrapped materials using flexible sheet-type dielectric barrier discharge. Applied Physics Letters. 2008; 93(22):221502. http://dx.doi.org/10.1063/1.3039808. 

Foegeding PM. Ozone inactivation of bacillus and clostridium spore populations and the importance of the spore coat to resistance. Food Microbiology. 1985; 2(2):123-34. http://dx.doi.org/10.1016/S0740-0020(85)80005-8. 

Fuente J, Bix L. Medical device packaging. In: Yam KL. The Wiley encyclopedia of packaging technology. Hoboken: John Wiley & Sons; 2009. p. 713-27. 

Govindaraj S, Muthuraman MS. Systematic review on sterilization methods of implants and medical devices. Int J ChemTech Res. 2015; 8(Jun):897-911. 

Grootveld M, Baysan A, Siddiqui N, Sim J, Silwood C. History of the clinical applications of ozone. Ozone: The revolution in dentistry. London: Quintessence Publishing; 2004. 

Hunt NK, Marinas BJ. Kinetics of escherichia coli inactivation with ozone. Water Research. 1997; 31(6):1355-62. http://dx.doi.org/10.1016/S0043-1354(96)00394-6.

International Organization for Standardization. BS EN ISO 11607-1:2006 packaging for terminally sterilized medical devices. Part 1: requeriments for materials, sterile barrier systems and packaging systems. Geneva: ISO; 2006. 

Iwamura T, Nagano K, Nogami T, Matsuki N, Kosaka N, Shintani H, Katoh M. Confirmation of the sterilization effect using a high concentration of ozone gas for the bio-clean room. Biocontrol Science. 2013; 18(1):9-20. PMid:23538847. http://dx.doi.org/10.4265/bio.18.9. 

Killeen S, McCourt M. Decontamination and sterilization. Surgery. 2012; 30(12):687-92. http://dx.doi.org/10.1016/j.mpsur.2012.10.008. 

Kogelschatz U. Dielectric-barrier discharges: their history, discharge physics, and industrial applications. Plasma Chemistry and Plasma Processing. 2003; 23(1):1-46. http://dx.doi.org/10.1023/A:1022470901385. 

Luqueta GR, Santos ED, Pessoa RS, Maciel HS. Wireless sensor network for ozone sterilization monitoring. Revista IEEE América Latina. 2016; 14(5):2167-74. http://dx.doi.org/10.1109/TLA.2016.7530410. 

Mahfoudh A, Moisan M, Séguin J, Barbeau J, Kabouzi Y, Kéroack D. Inactivation of vegetative and sporulated bacteria by dry gaseous ozone. Ozone Science and Engineering. 2010; 32(3):180-98. http://dx.doi.org/10.1080/01919511003791971. 

Mitsugi F, Nagatomo T, Takigawa K, Sakai T, Ikegami T, Nagahama K, Ebihara K, Sung T, Teii S. Properties of soil treated with ozone generated by surface discharge. IEEE Transactions on Plasma Science. 2014; 42(12):3706-11. http://dx.doi.org/10.1109/TPS.2014.2350533. 

Ohkawa H, Akitsu T, Ohtsuki K, Ohnishi M, Tsuji M. High-grade disinfection using high-density ozone. Journal of Advanced Oxidation Technologies. 2004; 7(2):154-60. 

Ozen BF, Mauer LJ, Floros JD. Effects of ozone exposure on the structural, mechanical and barrier properties of select plastic packaging films. Packaging Technology and Science. 2002; 15(6):301-11. http://dx.doi.org/10.1002/pts.602. 

Peeters FJJ, Rumphorst RF, Van De Sanden MCM. Dielectric barrier discharges : evidence for mobile surface charge. Plasma Sources Science & Technology. 2016; 25(3):1-6. http://dx.doi.org/10.1088/0963-0252/25/3/03LT03. 

Polovka M, Polovková J, Vizárová K, Kirschnerová S, Bieliková L, Vrška M. The application of FTIR spectroscopy on characterization of paper samples, modified by Bookkeeper process. Vibrational Spectroscopy. 2006; 41(1):112-7. http://dx.doi.org/10.1016/j.vibspec.l.01.010. 

Rice RG. Applications of ozone for industrial wastewater treatment: a review. Ozone Science and Engineering. 2012; 2010(18):37-41. 

Rickloff JR. An evaluation of the sporicidal activity of ozone. Applied and Environmental Microbiology. 1987; 53(4):683-6. PMid:3579276. 

Rutala WA, Weber DJ. Current principles and practices: new research; and new technologies in disinfection, sterilization, and antisepsis. American Journal of Infection Control. 2013; 41(5 Suppl):S1. PMid:23622739. http://dx.doi.org/10.1016/j.ajic.2012.12.003. 

Rutala WA, Weber DJ. Disinfection, sterilization, and antisepsis: an overview. American Journal of Infection Control. 2016; 44(5 Suppl):e1-6. PMid:27131128. http://dx.doi.org/10.1016/j.ajic.2015.10.038. 

Sakurai M, Takahashi R, Fukunaga S, Shiomi S, Kazuma K, Shintani H. Several factors affecting ozone gas sterilization. Biocontrol Science. 2003; 8(2):69-76. http://dx.doi.org/10.4265/bio.8.69. 

Samaranayake WJM, Miyahara Y, Namihira T, Katsuki S, Hackam R, Akiyama H. Ozone production using pulsed dielectric barrier discharge in oxygen. IEEE Transactions on Dielectrics and Electrical Insulation. 2000; 7(6):849-54. http://dx.doi.org/10.1109/94.891999. 

Seavey R. Packaging for sterilization: it’s not just a box, a bag or gift wrapping! Manag Infect Control. 2008; 8:82-97. 

Seavey R. Can a rigid container system be greener and safer at the same time? Managing Infection Control. 2009; 9(11):90-4. 

Somalwar RS, Janekar VU, Umate B. Advance method for calculating ozone chamber parameters. IOSR Journal of Electrical and Electronics Engineering. 2014; 1:33-8. 

Srinivasan K, Chitra S. The application of ozone in dentistry: a systematic review of literature. Scholar Journal of Dental Sciences. 2015; 2(6):373-7. 

Varga L, Szigeti J. Use of ozone in the dairy industry: a review. International Journal of Dairy Technology. 2016; 69(2):157-68. http://dx.doi.org/10.1111/1471-0307.12302. 

Walzak MJ, Flynn S, Foerch R, Hill JM, Karbashewski E, Lin A, Strobel M. Uv and ozone treatment of polypropylene and poly(ethylene-terephthalate). Journal of Adhesion Science and Technology. 1995; 9(9):1229-48. http://dx.doi.org/10.1163/156856195X01012. 

World Health Organization. Medical device regulations global overview and guiding principles. Journal of Healthcare Information Management. 2003; (3):36-40. 

Von Woedtke T, Kramer A. The limits of sterility assurance. GMS Krankenhaushygiene Interdisziplinär. 2008; 3(3):1-10. PMid:20204091.

59e9ef250e8825fd709aaadf rbejournal Articles
Links & Downloads

Res. Biomed. Eng.

Share this page
Page Sections