USO DE DUPLA PLACA NA ESTABILIZAÇÂO DE OSTEOTOMITAS EM OSSO CORTICAL LONGO - ANÁLISE DASD DEFORMAÇÕES "IN VITRO" COM "STRAIN GAUGES" E EXPERIMENTO "IN VIVO" EM CARNEIROS

USE OF DOUBLE PLATES IN THE ESTABILIZATION OF OSTEOTOMIES IN LONG CORTICAL BONE - ANALYSIS OF DEFORMATIONS "IN VITRO" WITH "STRAIN GAUGES" AND EXPERIMENTS "IN VIVO" IN SHEEPS

Shimano, Antonio Carlos; Paulin, J.B.P.; Paccola, C.A.J.

Downloads: 0
Views: 685

Resumo

Em fraturas tratadas com placa onde não há apoio na cortical oposta pela presença de cominuição, ou pela falta de compressão, há frequentes fraturas da placa neste nível. Para solucionar este problema é sugerido adicionar uma segunda placa de polietileno de alta densidade, com caracteristicas dimensionais determinadas, para servir como apoio. Os ensaios 11 "in vitro" foram realizados em tíbias de carneiros, utilizando "Strain Gauges" colocados nas placas. Os ensaios "in vivo" foram feitos em carneiros, com cirurgias bilaterais. A avaliação dos resultados foi feita através de experimentos "in vitro" e "in vivo". O experimento "in vivo" foi durante o período de 8 semanas. As conclusões mostraram a viabilidade do uso de dupla placa, salientando as vantagens clínicas e mecânicas com relação ao método clássico de osteossíntese com única placa e enxerto, utilizados nas fraturas propostas.

Palavras-chave

.

Abstract

There are comminutted fractures in which it is impossible to achieve apposite cortical contact after plating. When this happens there are often plate fracture (at the level of the fracture) or loosening of the implants in one or both bone segments. We are looking to solve this problem in adding a second plate to the opposite cortex. Elasticity of the second plate and its design are very and weakness of the bone underneath the two plates. We performed theoretical calculations and compression tests in several tests specimens "in vitro" for the predimensioning the plate. We placed " strain gauges " in the plates sewed to sheep tibie to register the deformation and the load. The " in vivo" tests were done in sheeps, on both legs, one of them was used as control ( conventional plating and bone grafting tothe apposite cortex ). The results were based on the analysis of the X-rays and compression mechanic tests. The use of double plate maybe acceptable for the clinical use, bringing evident advantages as compared to just ane plate.
5889fbbe5d01231a018b4790 ceb Articles
Links & Downloads

Res. Biomed. Eng.

Share this page
Page Sections