ELEVADORES ANGULADOS DE SELDIN. AVALlAÇAO ANALITICA DOS FATORES INTERFERENCIAIS PRESENTES NA FORMA E DIMENSOES DAS SUAS LAMINAS, SOBRE A FORÇA FINAL DESENVOLVIDA.

Seldin angled elevators. Analitical evaluation of the intervening factors presented in the shape and in the dimensions of its blades, along the fi nal developed force.

Campos, A.C. de; Gregori, C.; Tachibana, T.I.; Campos, T.N.

Downloads: 0
Views: 747

Resumo

Foram utilizados 26 elevadores dentais, previamente usina dos, de números IR e lL. As mensuraçoes das dimensões dos elevado res foram obtidas do traçado geométrico no desenho da projeção ob tida de cada elevador, sobre um papel vegetal, aposto na tela do projetor de perfil e usando-se uma máquina de medir tridimensional "Micro-Cord". Cada elevador desalojou a raiz dentária de resina de um modelo do segmento posterior da arcada dentária mandibular, tam bém em resina. As raízes foram fixadas aos modelos com um selante ã base de silicone. Posicionamos a ponta da lâmina do elevador no espaço exis tente entre a raiz de resina e o modelo suporte, e acionamos a ma nivela do DEE, imprimindo um movimento de rotação na lâmina do ele vador, resultando no desalojamento da raiz. Uma célula de carga me diu a força desenvolvida para o desalojamento da raiz, atravês d~ um sinal elétrico, transmitido ao registrador potenciomêtrico. O regisrro gráfico da força permitiu-nos calcular a força máxima. A análise dos resultados obtidos pela metodologia utiliza da, que analisa a relação entre as diferentes dimensões das lâmi nas de elevadores angulados de Seldin com as variações da força m~ xima presente na ponta da lâmina, no instante em que suplanta uma resistência padronizada, permite concluir que existem relações aI tamente significantes para as variáveis Comp. LC, raio externo (De) e o ângulo (6). Se estas relações forem consideradas em forma conjunta, as únicas variáveis influentes são o comprimento LC e o raio extern~ (pe).

Palavras-chave

Instrumentos odontológicos, Biomecânica, Engenharia biomedica

Abstract

Twenty-six different elevators (n9 IR and lL), which were previously prepared, were used. The physical dimen~ions of the eleva tors were obtained from the tracing of the projection of the eleva tors made through the use of a profile projector and the"Micro-cord" three-dimensional measuring device. Each elevator dislodged a resin- made dental root from a model of the posterior segment of the mandi bular dental arch, which was also made of acrylic. The rDots were fastened to th~ model jaw with a silicone-based sealant. The tip of the elevator blade was placed on the space bet ween the resin-root and the supporting model jaw. The DEE was then activated, forcing the rotational movement of the elevator tip, re sulting in the dislodging of the acrylic root. A measuring cell, mea sured the force which dislodged the root. This measurement was made the transmission of an electronic signal to a potentiomete·r. A gr~ phic representation of the force was used to determine the maximum force. The analysis of the results obtained from the methodology used in the atudy of the relation of the different physical dimensi ons of the Seldin angled elevator blades and the variations of the maximum mechanical force present at the tip of the blades when a standard resistance is exceeded led us to conclude that there are highly significant relations for comp. LC (central line lenght), out sjde radius (pe) and angle (/3) variables. However, if these relati ons are considered jointly, the only significant values are LC (central line length) and outside radius (pe).

Keywords

Dental instruments; Biomechanics; Biomedical engeneering